Containment Hoods - Hepa Filter Systems by SAS

SAS Certification Standards
Model# SS-330-DCH and SS-300-WSL

I. Velocity:
    Maximum 175 FPM Average Inlet Velocity on High Speed, with Clean Filters.
Minimum 125 FPM Average Inlet Velocity on Low Speed
II. Sound Levels:
    High - 66 dba
Low - 61 dba
III. Leakage:
    The fan cabinet interior is painted and sealed to minimize the potential for particulate bypass. The filter chamber has an active pressurized sealing mechanism. A single piece continuous closed cell neoprene gasket without seams is utilized on the filters to maximize seal integrity.
IV. Filter Efficiency:
    This unit is furnished with a HEPA filter having an efficiency of up to 99.97% on particles down to .3 microns in size.
    This unit can also be equipped with an optional ULPA filter having an efficiency of up to 99.9995% on particles down to .12 microns in size.


Note:  This product is intended to reduce operator exposure during compounding operations. We cannot guarantee that using this product will entirely eliminate such exposure.


Sentry Air Systems, Inc.
6999 West Little York, Ste. P1
Houston, TX.  77040

USA: (800) 799- 4609

Intl: (713) 690- 2153

Fax: (713) 690- 7872
Email: sales@sentryair.com

[Home]   [Products]   [Videos]   [Data]   [Contact Us]

©Sentry Air Systems, Inc. 2001 All rights reserved